Limitations
What the lesion detectors cannot see
This is the most important section on the site. Our classical detectors find microaneurysms, haemorrhages and hard exudates. They do not find the following — and in every case the system reports unknown rather than absent, because treating "I cannot see it" as "it is not there" would systematically under-grade the sickest patients.
Neovascularisation
New vessel growth is what defines proliferative disease — ICDR grade 4, the most sight-threatening grade. We have no detector for it.
Venous beading
One of the three limbs of the 4-2-1 rule for severe non-proliferative disease.
IRMA
Intraretinal microvascular abnormalities — the third limb of the 4-2-1 rule.
Deliberately not built
| Not built | Why |
|---|---|
| A trained lesion segmentation network | IDRiD provides 54 pixel-annotated training images. A network trained on that is a model of 54 images, not of diabetic retinopathy. We use classical morphology instead, which needs no training data and is inspectable. |
| Messidor-2 validation | Access approval takes days. IDRiD's test set is the external holdout instead. |
| Five-fold cross-validation | A single stratified split with a fixed seed. Cross-validation would give tighter error bars and buys nothing else here. |
| Authentication and patient records | This is a screening demonstrator, not a clinical system. There is no login and no patient database, and no real patient data should be uploaded to it. |
What has not been validated
- No field trial on a real portable fundus camera. The quality gate exists because portable cameras produce worse images, but we have not put one in front of it. That is the next step, not a claim.
- Quality-gate thresholds are provisional until they are fitted against a few hundred human-labelled images. Until then every response carries
thresholds_fitted: falseand the report page says so. - Unaided read time is an assumption. The staffing saving in the workflow model depends on how long an ophthalmologist takes to grade a raw photograph, which we have not measured. It is reported as a range across a sensitivity sweep rather than a single number.
- Vessel segmentation is a supporting component, not a leaderboard entry. It exists to stop vessels being counted as haemorrhages.
How the numbers are guarded
The dashboard reads its figures from a results file. If an evaluation has not been run, it prints not run yet — there is no placeholder value anywhere that could be shipped by accident. Validation figures are labelled as tuned-on and distinguished from the external holdout, which is opened once against a frozen configuration; the holdout runner refuses a second look unless explicitly overridden, and records it permanently if you do.